You are in:
/ Hettich Haematokrit 210

Hettich Haematokrit 210

Hettich Haematokrit 210 centrifugeWith this centrifuge, safety is top priority.

At 16,060 g sedimentation is completed in just a little over 6 minutes.  At any rate, a run time of 10 minutes is sufficient.  The rotor comes to a braked standstill.  Haematocrit determination with standard capillaries is carried out with 24-place rotor no. 2076.  Each capillary is secured within its own chamber and supported by a holding tray.  This holding tray serves to contain glass shards and any leakage should a tube break.  It is easy to clean and disposable in the case of glass breakage.  The rotor 2076′s lid is cover and evaluation disk in one.  Capillaries,  to be used for quantitative buffy coat analysis are centrifuged in 20-place rotor no. 2056 prior to further analysis.

Haematocrit centrifuge, without rotorHaematokrit 210
Power Supply208 - 240V
Frequency50-60Hz
Consumption250 VA
Emission, ImmunityEN / IEC 61326-1, class B
Max. Capacity20 / 24 capillaries
Max. RCF16,060
Max. RPM13,000 min
Running Time1 - 99 min, ∞ continuous run, short cycle mode (impulse key)
Dimensions (H x W x D)247 x 275 x 330 mm
Weightapprox. 10 kg

Product Code

  • SA(2104) – Hettich Haematokrit 210 centrifuge.

Click here for Hettich Haematokrit 210 accessories

Click here for Hettich Haematokrit 210 resources library information